Can 2 μL pipettes be inspected? [Pipette Accuracy Testers]

A volume of 2 μL is about 2 mg and the balance requires a resolution of 1 μg (0.001 mg). The AD4212B-PT can display mass at a resolution of 0.001 mg up to 5 g so 2 μL pipettes can be inspected. One μg is about 1,000,000th of a one yen coin (about 2,500,000th of an American penny). As you can imagine, measuring at 1 μg is not a simple task and the measuring environment is extremely important. When setting up the tester, ensure that the following conditions are met.

  • Avoid measuring on a weak foundation, on the second floor or higher, in the center of a floor away from the support pillars, in earthquake-proof structures, near tall structures, along heavily traveled roadways, etc.
  • Place the balance on a stand.
  • Prevent direct contact with air blown from air conditioners, etc.
  • Perform tests in a temperature and humidity controlled room.
If these conditions are met and measurement values are still not stable, it is necessary to take further countermeasures, such as decreasing the precision level of the values displayed by the balance. (For example, by pressing SAMPLE to select a resolution of 0.01 mg.)