I wish to send weighing data at certain intervals to a PC using real-time transfer mode. [Weighing Environment Logger] [AD-1687]

Can only be used if the balance has "interval output mode". (Models: FX-i/FZ-i, BM, HR-AZ/A)

  1. Using the balance's instruction manual, turn on "interval output mode".
  2. Change the "interval time" settings to match the interval at which you wish to record.
    Note: The balance's "interval time" should be set to at least 10 seconds. If the interval is too short, saved data may be left out depending on the PC's performance.
  3. Insert the balance's RS-232C cable into the AD-1687 while the power is turned off.
  4. Insert a USB cable into the AD-1687, select "REAL TIME" and hit the ENTER key.
  5. Place the cursor on the application you would like to save to (EXCEL etc.)
  6. Initiate interval output from the balance (example: press the PRINT key on the balance)
    According to the interval time settings for the balance, measurement data and environmental data will be transmitted to the application.