Does it affect the measurement result if the immersion level of the sensor plates change? [Viscometers] [SV10/100]
Although it depends on viscosity of each sample fluid, if the level changes by 1mm with SV-10 in low viscosity range, measurement error of approx. 2-3%, and with SV-100 in high viscosity range, approx. 10-15%, could be produced.